Executing Self-Test and Leak Test
Executing Self-Test
Self-Test Programming Example
This sample program is effective for an E5250A that has four cards installed. If your
E5250A does not have four cards installed, delete the Relay Test program lines for
the empty card slots. The program runs on HP BASIC.
10
! Self-Test by External Controller
20
!
30
REAL A
40
ASSIGN @Hp5250 TO 722
50
OUTPUT @Hp5250;":DIAG:TEST:FRAM:CLE CONT"
60
OUTPUT @Hp5250;":DIAG:TEST:FRAM:CLE FPAN"
70
OUTPUT @Hp5250;":DIAG:TEST:CARD:CLE ALL"
80
DISP "Press Continue to go to Controller Test."
90
PAUSE
100 ! Controller Test
110
OUTPUT @Hp5250;":DIAG:TEST:FRAM:EXEC? CONT"
120
ENTER @Hp5250;A
130
IF A=0 THEN GOTO 160
140
DISP "Failed. Press Continue to go to I/F Test."
150
GOTO 170
160
DISP "Passed. Press Continue to go to I/F Test."
170
PAUSE
180 ! Front Panel I/F Test
190
OUTPUT @Hp5250;":DIAG:TEST:FRAM:EXEC? FPAN"
200
DISP "Press Local/Self Test Key within 10 sec."
210
ENTER @Hp5250;A
220
IF A=0 THEN GOTO 250
230
DISP "Failed. Press Continue to go to Relay Test."
240
GOTO 260
250
DISP "Passed. Press Continue to go to Relay Test."
260
PAUSE
270 ! Relay Test
280
OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 1"
290
ENTER @Hp5250;A
300
IF A=0 THEN GOTO 330
310
DISP "Card 1 failed. Press Continue to test Card 2"
320
GOTO 340
330
DISP "Card 1 passed. Press Continue to test Card 2"
340
PAUSE
350
OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 2"
360
ENTER @Hp5250;A
370
IF A=0 THEN GOTO 400
380
DISP "Card 2 failed. Press Continue to test Card 3"
390
GOTO 410
400
DISP "Card 2 passed. Press Continue to test Card 3"
410
PAUSE
420
OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 3"
3-8