Softkey Menu 2:
For the following conditions, the following softkeys are displayed by selecting the
MORE softkey on Softkey Menu 1.
Cursor points to row of normal input port or to row of input Bias Port with Bias
Mode OFF.
Softkey
OPEN
Opens all relays. If Bias Mode is ON, then closes relays to connect
ALL
bias enabled output ports to the input Bias Port.
OPEN
Opens all relays of row at cursor. If Bias Mode is ON, then closes
ROW
relays to connect bias enabled output ports (that are not connected
to other input ports) to the input Bias Port.
CLOSE
Closes all relays of row at cursor. And all output ports are
ROW
disconnected from the input Bias Port.
MORE
Displays softkey menu 1.
RETURN
Displays the previous level of softkey menu.
Softkey Menu 3:
For the following conditions, the following softkeys are displayed by selecting the
MORE softkey on Softkey Menu 1:
Cursor points to row of input Bias Port and Bias Mode is ON.
Softkey
OPEN
Opens all relays, then connects all bias enabled output ports to the
ALL
input Bias Port.
DISABLE
Bias disables all output ports. So relays open to disconnect all
ROW
output ports from the input Bias Port.
ENABLE
Bias enables all output ports, then connects all output ports (that
ROW
are not connected to another input port) to the input Bias Port.
MORE
Displays softkey menu 1.
RETURN
Displays the previous level of softkey menu.
Agilent E5250A User's Guide, Edition 11
Controlling the E5250A
Using Virtual Front Panel Utility
Description
Description
5-41