ENTER @Hp5250;A
440
IF A=0 THEN GOTO 470
450
DISP "Card 3 failed. Press Continue to test Card 4"
460
GOTO 480
470
DISP "Card 3 passed. Press Continue to test Card 4"
480
PAUSE
490
OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 4"
500
ENTER @Hp5250;A
510
IF A=0 THEN GOTO 540
520
DISP "Card 4 failed."
530
GOTO 550
540
DISP "Card 4 passed."
550
END
Line Number
30 to 90
Clears all previous test results.
100 to 170
Executes the Controller Test, then returns test result.
180 to 260
Executes the Front Panel Interface Test, then returns the test
result.
270 to 340
Executes the Relay Test for the card installed in slot 1 of the
E5250A, then returns test result.
350 to 410
Executes the Relay Test for the card installed in slot 2 of the
E5250A, then returns test result.
420 to 480
Executes the Relay Test for the card installed in slot 3 of the
E5250A, then returns test result.
490 to 550
Executes the Relay Test for the card installed in slot 4 of the
E5250A, then returns test result.