Agilent Technologies Agilent E5250A User Manual page 274

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

Table 9-5
NOTE
Executing Sample Programs
HCI Measurement Program
HCI Degradation Measurement Flow
step
1
Initializes the Instruments:
1. Resets the 4155/4156.
2. Resets and initializes the E5250A.
Configuration Mode: AUTO
Bias Mode: OFF
Connection Rule: FREE
Channel Status: All Relays Open
2
Determines Stress Condition. See later.
3
Selects devices valid for the HCI test. See later.
4
Characterizes the initial parameters. See later.
5
Applies stress and characterizes the parameters.
See later.
Step 2. Determines Stress Condition
This step determines the gate stress voltage (Vgstr) from the Ib-Vgs measurement
test result. This Vgstr value will be used in step 5.
Test Device for Defining Stress Conditions
The test device used to determine the stress conditions (step 2) should not be used
for Hot-Carrier stress testing (steps 3 to 5).
9-18
Description
Agilent E5250A User's Guide, Edition 11
4155/4156
setup file
IBVG.MES
IGLEAK.MES
IDLEAK.MES
PARAM.MES
PARAM.MES