NOTE
Figure 3-1
NOTE
If Fail LED is ON after Relay Test
After executing the relay test, if the Fail LED turns on, the cause is one of the
following:
• The Relay Test Adapter is not connected or a cable is still connected to input
connector.
• Output is not open. See step 2 of previous procedure.
• You did not properly install the E5255A. See "To Configure E5255A" on page
2-10.
• A plug-in card may be defective. Contact your nearest Agilent Technologies
Service Center.
Relay Test Adapter Connection
Settings after Relay Test
After Relay Test, the E5250A settings are changed to same as after :SYST:CPON
command execution.
Agilent E5250A User's Guide, Edition 11
Executing Self-Test and Leak Test
Executing Self-Test
Relay Test Adapter
3-5