Agilent Technologies Agilent E5250A User Manual page 156

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

Program List:
Programming the E5250A
Programming Examples
10
! Example: Using Couple Port
20
!
30
INTEGER Complete
40
DIM Channel$[100]
50
! (Substrate,Gate,Drain,Source)
60
Channel$="(@101,202,303,505)"
70
!
80
! Put Instrument Initialize Routine here
90
!
100 ASSIGN @Hp5250 TO 722
110 OUTPUT @Hp5250;"*RST"
120 OUTPUT @Hp5250;":ROUT:FUNC ACON"
130 OUTPUT @Hp5250;":ROUT:CONN:RULE ALL,SROU"
140 OUTPUT @Hp5250;":ROUT:CONN:SEQ ALL,BBM"
Line
Number
40
Declares data size of Channel$, which is used as channel_list
parameter.
60
Defines the desired channel_list.
80
Here, insert the initialization routine for your measurement
instrument.
110
Resets the E5250A.
120
Sets the E5250A to Auto channel configuration mode.
130
Sets the connection rule to Single.
140
Sets the connection sequence to Break-Before-Make.
6-20
Description
Agilent E5250A User's Guide, Edition 11