Syntax
Example
Syntax
Example
Command Reference
[:ROUTe]:BIAS:CHANnel:DISable:CARD
[:ROUTe]:BIAS:CHANnel:DISable:CARD
This command bias disables all output ports of the specified card.
When Bias Mode is ON, this command opens relays to disconnect the input Bias
Port from all output ports. Refer to "ROUTe:BIAS:STATe".
At *RST, all output ports are bias enabled.
[:ROUTe]:BIAS:CHANnel:DISable:CARD card_number
Parameter
card_number
OUTPUT @Hp5250;":ROUT:BIAS:CHAN:DIS:CARD ALL"
[:ROUTe]:BIAS:CHANnel:DISable[:LIST]
This command bias disables the output ports specified by the channel_list.
When Bias Mode is ON, this command opens relays to disconnect the input Bias
Port from the output ports specified by channel_list. Refer to
"ROUTe:BIAS:STATe".
At *RST, all output ports are bias enabled.
[:ROUTe]:BIAS:CHANnel:DISable[:LIST] (@channel_list)
Parameter
channel_list
The input ports specified by the channel_list are ignored. The input port is always
the input Bias Port. So, in the channel_list, only the card numbers and output ports
have meaning. But you cannot omit the input port in the channel_list.
OUTPUT @Hp5250;":ROUT:BIAS:CHAN:DIS (@10301,10402)"
In above example, the input Bias Port on card 1 is disabled for output ports 1 and 2.
Specified input ports 3 and 4 are ignored because the input port is always the input
Bias Port.
7-22
Explanation
card number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
Explanation
Channels to bias disable.
For details about channel_list, refer to Chapter 5.
Agilent E5250A User's Guide, Edition 11