Agilent Technologies Agilent E5250A User Manual page 231

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

Syntax
Parameters
Syntax
Parameters
hpe5250a_biasChanCard
This function will enable or disable bias on all the output ports of the specified card.
ViStatus _VI_FUNC hpe5250a_biasChanCard(ViSession vi,
ViInt16 disable_enable, ViInt16 bias_cardno);
vi
Instrument handle returned from hpe5250a_init( ).
disable_enable
Bias status.
0 : sets bias enabled card.
1 : sets bias disabled card.
bias_cardno
Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5
(all card) in the normal configuration mode, or 0 (all card in the
automatic configuration mode). For the configuration mode, see
hpe5250a_func.
hpe5250a_biasChanList
This function will enable or disable bias on all the output ports specified by the
biaschan_list.
The parameter 'biaschan_list' is an array of integers with each integer representing
one channel. The last number of the 'biaschan_list' should be "0" (numeric zero) to
identify the end of the list. The maximum number of channels that can be specified
by the list is 100.
ViStatus _VI_FUNC hpe5250a_biasChanList(ViSession vi,
ViInt16 biaschan_disen, ViInt32 _VI_FAR biaschan_list[ ] );
vi
Instrument handle returned from hpe5250a_init( ).
biaschan_disen
Bias status.
0 : sets bias enabled port.
1 : sets bias disabled port.
biaschan_list[ ]
Channel numbers. 5 digits integer. ABCDE. where A: card
number, BC: input port number, DE: output port number.
Top zero(s) can be ignored. For example, if A=0, BC=01, and
DE=01, channel number should be 101 instead of 00101.
Agilent E5250A User's Guide, Edition 11