NOTE
Executing the Controller Test
To execute the controller test, press the CONTROLLER softkey.
If [F] result is displayed after the test, the E5250A is defective. Contact your
nearest Agilent Technologies Service Center.
Executing the Front Panel Interface Test
To execute the front panel interface test, do the following:
1. Press the FRONT PANEL softkey.
2. Confirm that the following four LEDs on the E5250A front panel are blinking.
And press the
Local/Self Test
Remote LED
System Fail LED
Fail LED
Local/Self Test LED
Fail Condition for Front Panel Interface Test
If an LED does not blink after selecting the FRONT PANEL softkey, the LED is
defective.
If you do not press the
If you pressed
Local/Self Test
it means the
Local/Self Test
Contact your nearest Agilent Technologies Service Center.
Agilent E5250A User's Guide, Edition 11
key on the E5250A front panel within 10 seconds.
key within 10 seconds, the Fail LED turns on.
Local/Self Test
key within 10 seconds, and the Fail LED still turns on,
key is defective.
Executing Self-Test and Leak Test
Using the Self-Test Utility
3-13