Syntax
Query response
NOTE
Example
Command Reference
:SYSTem:CDEScription?
:SYSTem:CDEScription?
This query command returns a description of the specified card.
This command has query form only.
:SYSTem:CDEScription? card_number
Parameter
card_number
"card_description" <newline><^END>
If designated card is not installed or initialize failed, this command returns
"No Card".
For E5252A in Auto Config mode, "card_description" response is one of
following:
"E5252A 10x12 Matrix Switch"
"E5252A 10x24 Matrix Switch"
"E5252A 10x36 Matrix Switch"
"E5252A 10x48 Matrix Switch"
For E5252A in Normal Config mode:
"E5252A 10x12 Matrix Switch"
For E5255A in Auto Config mode:
"E5255A I1,I2,I3,I4,I5,I6 Channel Multiplexer"
Where In is the number (two characters) of output ports for each input port n.
For example:
"E5255A 24,24, 0,16,16,16 Channel Multiplexer"
For E5255A in Normal Config mode:
"E5255A 24 (8x3) Channel Multiplexer"
OUTPUT @Hp5250;":SYST:CDES? 1"
ENTER @Hp5250;A$
7-44
Explanation
card number:
For Auto Config mode: 0
For Normal Config mode: 1, 2, 3, or 4
Agilent E5250A User's Guide, Edition 11