Table 6-1
Description
1
Defines Channel Configuration Mode
2
Defines Connection Rule
3
Defines Connection Sequence
4
Using Bias Mode
Selects input Bias Port (for E5252A)
Bias enables all output ports
Bias enables specified output ports
Bias disables all output ports
Bias disables specified output ports
Sets Bias Mode ON/OFF
5
Using Couple Port (for E5252A)
Sets up Couple Port
Sets Couple Port Mode ON/OFF
6
Controls Relays on Cards
CAUTION
Programming Summary
:ROUT:FUNC channel_config
:ROUT:CONN:RULE card_no, rule
:ROUT:CONN:SEQ card_no,sequence
:ROUT:BIAS:PORT card_no, input_port_no
:ROUT:BIAS:CHAN:ENAB:CARD card_no
:ROUT:BIAS:CHAN:ENAB:LIST channel_list
:ROUT:BIAS:CHAN:DISA:CARD card_no
:ROUT:BIAS:CHAN:DISA:LIST channel_list
:ROUT:BIAS:STAT card_no, state
:ROUT:COUP:PORT card_no, 'input_port_no'
:ROUT:COUP:STAT card_no, state
:ROUT:CLOS channel_list
:ROUT:OPEN channel_list
:ROUT:OPEN:CARD card_no
Parameter
channel_config
rule
sequence
state
channel_list
card_no
input_port_no
For details about these parameters, refer to Chapter 5 and Chapter 7.
Bias Port and Couple Port cannot be set for the same input port number.
Agilent E5250A User's Guide, Edition 11
Command
Description
NCON or ACON.
FREE or SROU.
NSEQ, BBM or MBBR.
ON (1) or OFF (0).
5 digit decimal number. See Chapter 5.
ALL or 0 or 1 to 4 are available.
1 to 10 are available for :ROUT:BIAS:PORT.
1, 3, 5, 7 or 9 are available for :ROUT:COUP:PORT.
Programming the E5250A
Creating a Control Program
Default Setting
NCON
ALL,FREE
ALL,BBM
ALL,10
ALL
All enabled
All enabled
All enabled
ALL,OFF
No couple ports
ALL,OFF
All relays open
All relays open
ALL
6-7