Agilent Technologies Agilent E5250A User Manual page 324

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

SCPI Command Summary
Command
[:ROUT]:CONN:SEQ? card_number
[:ROUT]:COUP:PORT card_number,'couple_port'
[:ROUT]:COUP:PORT? card_number
[:ROUT]:COUP[:STAT] card_number,state
12-6
Description
Returns the connections sequence mode of the
specified card:
NSEQ: No-Sequence mode
BBM: Break-Before-Make mode
MBBR: Make-Before-Break mode.
card_number:
For Auto Config mode: 0
For Normal Config mode: 1, 2, 3, 4
Sets up the couple ports for the specified card. Only
for the E5252A card.
card_number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
couple_port:
One or more input port numbers: 1, 3, 5, 7, or 9.
Enclose by single quotation marks. Separate
multiple input port numbers by comma. For
example: '1,5'
Returns the lower input port number of each couple
pair on the specified card. Only for the E5252A
card.
card_number:
For Auto Config mode: 0
For Normal Config mode: 1, 2, 3, 4
Turns Couple Mode ON or OFF for the specified
card. Only for the E5252A card.
card_number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
state:
ON or 1: turn ON Couple Mode
OFF or 0: turn OFF Couple Mode