2.5 Subscan Unit
7.
Remove the 4 fixing screws.
8.
Holding the cassette magnetizing plate as shown in the
right figure, and pull it out.
9.
Carry out reassembly in the reverse order to that described
above.
Remarks for installation
After confirming two positioning pins which are located
between the fixing screws are properly sit in the hole,
tighten the fixing screws.
Attach the anti-fall protection metal to its original position
from which it was detached in step 3.
Confirm that there is no foreign metals such as clip on
the magnetized surface of the cassette magnetizing
plate.
Adjustment after installation
Carry out
"3.6 Adjusting the Peel-Detection Rollers,
Carry out
"3.7 Adjusting the Pressing Amount,
Magnetizing Plate
p.117".
p.122".