2.5 Subscan Unit
7.
Remove the 4 fixing screws.
8.
Holding the cassette magnetizing plate as shown in the
right figure, and pull it out.
9.
Carry out reassembly in the reverse order to that described
above.
•
Remarks for installation
•
After confirming two positioning pins which are located
between the fixing screws are properly sit in the hole,
tighten the fixing screws.
•
Attach the anti-fall protection metal to its original position
from which it was detached in step 3.
•
Confirm that there is no foreign metals such as clip on
the magnetized surface of the cassette magnetizing
plate.
•
Adjustment after installation
Carry out
"3.6 Adjusting the Peel-Detection Rollers,
Carry out
"3.7 Adjusting the Pressing Amount,
Magnetizing Plate
p.117".
p.122".
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Fixing Screws