2.3 Insertion / Ejection Unit
12.
Unplug a connector, and remove the wire bundle from the
wire clamp.
JJ49
13.
Remove three fixing screws, then pull out the gear for the
ejection motor(M3) from the screw hole, and remove the
ejection motor(M3).
14.
Carry out reassembly in the reverse order to that described
above.
Remarks for Installation
Assembling the ejection shutter (step9.)
Confirm that the arm for shutter spring is properly set in the
damper gear assy.
Lubricate the grease inside the shutter spring.
Recommended grease : Epinox AP-2
First, insert the shutter shaft (right) in the bearing (right), then
insert the shutter shaft (left) in the bearing (left).
Fixing Screws (3pcs)
Ejection Motor (M3)
Shutter Shaft (Left)
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Connector
Shutter Spring
Damper Gear
Shutter Shaft (Right)