Konica Minolta REGIUS 190 Service Manual page 150

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

10.
With the steel scale, measure the gap between the floating
rail on the right side of the device and the absorption plate.
Criterion : 12 ±0.5mm
If the measured values does not satisfy the rating
Adjust the correction amount (mm) by turning the rotary switch
(SW4) on the MCB2.
No. Degree of
correction
(Pressing
amount)
0
+4.0
1
+4.5
2
+5.0
3
+5.5
4
-2.0
5
-1.5
6
-1.0
7
-0.5
I mp o r t a n t
When the MCB2 is changed ("2.7.1
Replacement of MCB2 (Mechanical Con-
trol Board)", p.81), set the correction
value for the new MCB2 as that of the old
one.
No. Degree of
correction
(Pressing
amount)
8
0
9
+0.5
A
+1.0
B
+1.5
C
+2.0
D
+2.5
E
+3.0
F
+3.5
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.7 Adjusting the Pressing Amount
Floating rail
Left side
MCB2
SW4
123