Konica Minolta REGIUS 190 Service Manual page 186

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

Error
Error name
code
GetStatus receipt
04208
time-out - front door
open
GetStatus receipt
04209
time-out - front door
close
04210
GetStatus receipt
time-out - error
GetStatus receipt
04211
time-out - Standby
GetStatus receipt
04212
time-out - transporting
in erase mode
GetStatus receipt
04213
time-out - power OFF
after Ready
RegistrationOK receipt
04214
time-out
Error description
Front door open Communication
for the purpose of reporting that
the front door is open to the SCB2
was not possible.
Front door closed Communication
for the purpose of reporting that
the front door is closed to the
SCB2 was not possible.
Error Communication for the pur-
pose of reporting occurrence of
an error to the SCB2 was not pos-
sible.
Standby Communication for the
purpose of reporting transition to
StandBy condition to the SCB2
was not possible.
Erasure-mode feeding Communi-
cation for the purpose of reporting
feeding of a cassette in Erasure
mode to the SCB2 was not possi-
ble.
Power Off after Ready Communi-
cation for the purpose of reporting
ejection of a cassette to the SCB2
was not possible.
Reception of either Registra-
tionOK or NoRegistration was not
possible within 10 seconds after
bar code detection.
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Action
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.