•
Important adjustment issues
Install the peel detect roller jig on the cassette magnetizing
plate, and adjust the standard position of the peel detect roller
and the position where the peel detect sensor switches to
"ON".
• When the verification part (plate thickness : 11.1mm)
passes through, it should not contact the peel detect roller.
• When the thinner part (plate thickness : 11.4mm) passes
through, though it is allowed to contact the peel detect roller,
but the peel detect sensor should remain OFF.
• When the thicker part (plate thickness : 11.6mm) passes
through, the peel detection sensor should switch to "ON".
•
Timing for adjustment
Adjustment is to be carried out whenever the peel detection
sensor or the absorption plate is replaced.
Peel Detection
Roller Test Jig
Magnetizing
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.6 Adjusting the Peel-Detection Rollers
Verification Part
(11.1mm)
Thinner Part (11.4mm)
Thicker Part (11.6mm)
Plate
119