Set Up Of Dummy Sbc2 Program; Installing The Dummy Sbc2 Program - Konica Minolta REGIUS 190 Service Manual

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

3.2 Set Up of dummy SBC2 program

To check each operation of mechanical control, use the dummy SBC2 program. Start the dummy SBC2 program
following the procedures below.
Im p o r t a n t
Do not change the setting of DIP SW
other than SW2- 1 on the MCB2. (for the
rest of DIP SWs, refer to ("
Setting, p.184"
I mp o r t a n t
To terminate the dummy SBC2 program,
confirm that the all dips of DIP SW (SW2)
are set to OFF (left).
3.2.1

Installing the dummy SBC2 program

Preparation of maintenance PC
Prepare a PC which satisfies the following specifications.
• OS: Windows2000 or later
• Interface: EIA-232C (D-Sub9P)
Installation
1.
Before start installation, check that all of following files are
available.
DmyScbEng.msi
InstMsiA.exe
InstMsiW.exe
setup.exe
setup.ini
2.
Start "setup.exe", and follow the messages shown on the
window.
When the installation completes successfully, a short-cut
to the dummy SBC2 program will be automatically cre-
ated.
Refer to
"A.1 SCB2 Dummy Program Screens, p.169"
details of dummy SBC2 program screens.
A.2 DIP Switch
)
for the
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01