3.2 Set Up of dummy SBC2 program
To check each operation of mechanical control, use the dummy SBC2 program. Start the dummy SBC2 program
following the procedures below.
Im p o r t a n t
Do not change the setting of DIP SW
other than SW2- 1 on the MCB2. (for the
rest of DIP SWs, refer to ("
Setting, p.184"
I mp o r t a n t
To terminate the dummy SBC2 program,
confirm that the all dips of DIP SW (SW2)
are set to OFF (left).
3.2.1
Installing the dummy SBC2 program
•
Preparation of maintenance PC
Prepare a PC which satisfies the following specifications.
• OS: Windows2000 or later
• Interface: EIA-232C (D-Sub9P)
•
Installation
1.
Before start installation, check that all of following files are
available.
•
DmyScbEng.msi
•
InstMsiA.exe
•
InstMsiW.exe
•
setup.exe
•
setup.ini
2.
Start "setup.exe", and follow the messages shown on the
window.
•
When the installation completes successfully, a short-cut
to the dummy SBC2 program will be automatically cre-
ated.
Refer to
"A.1 SCB2 Dummy Program Screens, p.169"
details of dummy SBC2 program screens.
A.2 DIP Switch
)
for the
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.2 Set Up of dummy SBC2 program