11.
After coupling the back plate with the front plate, transports
the cassette to the ejection position.
1)
Click [Absorb pos.] of "Transport Unit".
2)
Only when the cassette is mammo type, click [3mm up]
of "Sub scanning".
3)
Check that "Transport Unit Magnetizing Position ...OK"
is shown on "Information : " window.
4)
Click [DC motors] tab.
5)
Click [Lock] of "Lock-release motor".
6)
Check that "Lock...OK" is shown on "Information : " win-
dow.
7)
Click [Stepping motors] tab.
8)
Click [Eject pos.] of "Transport Unit".
9)
Check that "Transport Unit Ejection Position ...OK" is
shown on "Information : " window.
12.
Cassette ejection operation starts.
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.4 Verification of Reading Operation
113