2.3 Insertion / Ejection Unit
8.
Unscrew and remove the eight fixing screws* which are
securing the insertion/ejection unit in place. (*: Four on each
side)
9.
Remove the wire bundle from the clamp and pull the bundle
towards the front of the device.
Depending the position of the Transport Unit at the time,
it may require to tilt the Transport Unit using a handle so
that the hand can reach the part.
10.
Hold the insertion / ejection unit as shown in the right, and
remove from the main unit.
11.
Carry out reassembly in the reverse order to that described
above.
For reassembly, it will make it easier if you insert the unit
while aligning it to the left on the front of the main unit.