3.4 Verification of Reading Operation
9.
Subscanning operation starts.
1)
Check that the "Cassette back plate sensor" of "Sen-
sor" is "ON".
2)
Click [Sub scanning] of "Sub scanning".
Cassette magnetizing plate elevates.
Image reading by the optical unit will not be imple-
mented.
3)
Check that [V-SYNC] of "Sensor" switches from "OFF"
to "ON".
4)
Check that "Subscan : Subscanning ...OK" is shown on
"Information : " window.
10.
Erasing operation starts.
Erasing lamp will not be turned on.
1)
Click [Erase 1] of "Sub scanning".
2)
Check that "Subscan: Erase 1 ...OK" is shown on
"Information : " window.
112
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01