4.4 Error Code
Error
Error name
code
GoBackward receipt
04216
time-out
Laser ON w/o send-
04217
ing high voltage ON
(case 1)
Laser ON w/o send-
04218
ing high voltage ON
(case 2)
Illegal command
04219
received
160
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Error description
A GoBackward command (i.e.,
permission for erasure start after
reading) could not be received
from the SCB2 within 5 seconds
after the completion of subscan-
ning.
Even though a request for high
voltage On was not reported to
the SCB2, an attempt was made
to provide notification of Laser On
to the SCB2 (with no stand-by
cassette in the insertion slot).
Even though a request for high
voltage On was not reported to
the SCB2, an attempt was made
to provide notification of Laser On
to the SCB2 (with a stand-by cas-
sette in the insertion slot).
A command which was unrecog-
nizable to the MCB2 was received
from the SCB2.
Action
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
If this problem should reoccur, confirm
that the SCB2(CN5)-MCB2(CN4) con-
nectors are connected and that cables
are not damaged or broken.