Konica Minolta REGIUS 190 Service Manual page 22

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

Cautions Regarding Transpor-
tation and Storage
Cautions Before Inserting the
Cassette
Cautions Regarding Disposal
WARNING
• • • •
Temperature and humidity levels must be maintained within the following
specified ranges.
Before unsealing the
plate
Temperature
10:~40:
Humidity
Humidity 80%RH or lower (without condensation)
• • • •
Do not expose the plate to X or γ rays.
• • • •
If the cassette is transported or stored in its packing, ensure that the cassette
is kept in horizontal position. Storage of cassettes in vertical position for
extended periods or in a high-temperature environment may result in warping
of the back plate.
• • • •
Once the plate is unpacked, store it in the cassette.
• • • •
Do not store the front and back plates separately. In particular, storage of the
back plate separately in angled position may result in warping.
• • • •
Be careful not to drop or bump cassettes.
• • • •
As for previous cassettes, cassettes in use should be stored vertically in a
rack.
In order to avoid problems with cassette conveying operation inside the
REGIUS MODEL 190, the following cautions should be observed:
• • • •
Ensure that the locking mechanism is functioning properly and check for
bending or other damage. Do not insert damaged cassettes into the REGIUS
MODEL 190.
• • • •
Make sure that there is no paper clipped into the memo clip on the back
plate.
• • • •
Make sure that there are no foreign objects such as clips between the front
and back plates.
• • • •
In order to ensure that the back plate properly adheres to the
• • • •
absorption plate, check that there is nothing (e.g. tape) adhered to the back
plate.
• • • •
Ensure that the bar code has not become soiled.
Disposal must be carried out by following the local regulations in each
country.
• • • •
If the protective layer on the plate is damaged and the photo-stimulated fluo-
rescent coating leaking, be careful to handle the plate so that none of the
photo-stimulated fluorescent coating comes into contact with the body.
• • • •
Lead foil is incorporated into all plates (RP-3S) except plates for mammo-
grams (RP-3M). Disposal of such plates must be carried out in accordance
with the appropriate waste disposal regulations.
When disposing the REGIUS Plate, accept the regulations of each
local government.
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
After unsealing the plate
10:~30:
xv

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dd-941

Table of Contents