SW2-8
Setting Mode
ON
Nonexpanded
•
LCD Message Language
SW3-3
Setting Mode
OFF
Japanese
ON
English
• SW3-1, SW3-2, SW3-4, SW3-5, SW3-6, SW3-7, SW3-8 are spares.
•
Fine Adjustment of Magnetizing Position at Transport Unit.
SW4
Offset Value
0
+4.0mm
1
+4.5mm
2
+5.0mm
3
+5.5mm
4
-2.0mm
5
-1.5mm
6
-1.0mm
7
-0.5mm
8
0.0mm
9
+0.5mm
A
+1.0mm
B
+1.5mm
C
+2.0mm
D
+2.5mm
E
+3.0mm
F
+3.5mm
Description
Continuous operation without ejecting the cassette.
Description
Uses Japanese for the LCD message on the REGIUS 190.
Uses English for the LCD message on the REGIUS 190.
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
A.2 DIP Switch Setting
185