[Stepping Motors] Tab
Important
Full care should be taken when starting the stepping motor so that any component
parts should not interfere with each other.
Item / Key
Cassette
High resolution / Standard
Erase speed
Sub
[Initialize]
scanning
[Move to HP]
[Sub scanning]
[Erase1]
[Erase2]
[3mm down]
[3mm up]
Trans-
[Initialize]
port Unit
[Move to HP]
[Absorb pos.]
[Escape pos.]
[Eject pos.]
[Insertion pos.]
Description
Specifies the cassette type that will be used as a standard for operations
of subscan, Cassette receive Unit, justification.
Specifies the mode for subscanning.
Specifies the operation speed for erase (read mode).
Initiates initialization of the subscan unit.
Moves the subscan unit to HP.
Initiates the subscan operation (image reading operation) of the subscan
unit.
Initiates the erase operation (erase after image reading) of the subscan
unit.
Initiates the erase operation (erase in erase mode) of the subscan unit.
Descends the subscan unit by 3mm.
Ascends the subscan unit by 3mm.
Initiates initialization of the Transport Unit.
Moves the Transport Unit to HP.
Moves the Transport Unit to magnetizing position where the cassette is
magnetized onto the cassette magnetizing plate.
Moves the Transport Unit to magnetizing position where the cassette is
separated in to front and back plates.
Moves the Transport Unit to ejection position.
Moves the Transport Unit to insertion position.