4.1 Troubleshooting for Electrical Control & Communication
Case 3: Operation condition where incorrect settings have been made for the Reader, REGIUS Console,
or JM etc. Countermeasures involve the re-definition of IP and host name settings to ensure no
duplication.
After this, be sure to restart the system.
Cause
1) Two Readers exist with the
same IP and host name.
2) Two Readers exist with
the same IP.
3) Two REGIUS Consoles
exist with the same host
name.
4) Two JM exist with the
same IP and host name.
5) Two REGIUS Consoles
exist with the same IP.
6) Two JMs exist with the
same IP.
7) Two Readers exist with
the same host name.
8) Two JMs exist with the
same host name.
9) A Reader and REGIUS
Console share the same
IP and host name.
10) A Reader and JM share
the same IP and host
name.
11) A REGIUS Console and
JM share the same IP and
host name.
130
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Remedy
Between POWER ON and READY, both of the Readers will startup.
After one has performed reading, a "04214" error will occur on the other
when an attempt is made to insert a cassette, and following this, the same
error will occur for the first Reader.
If an attempt is made to simultaneously insert cassettes into both of these
Readers, a "13031" error will occur on one of these.
Although normal reading operations will be possible for one of the Readers,
after this operation has been started for the other, operation will halt as soon as
the LCD indicates "busy ?" and a "04202" error will occur.
If host names are duplicated on Windows 2000, a warning will be displayed, and
although the REGIUS Console will startup, the reading of images will not be
possible. Note however that the Reader will operate normally.
The first JM to startup will operate.
When host names are duplicated on Windows 2000, a warning will be dis-
played and reading will not be possible. During the first read, a certain
amount of time will be required between bar code detection and cassette
insertion.
If operation is attempted with only the JM started last, a registration error will
occur.
The REGIUS Console with the modified IP will not start and a warning will be
displayed on Windows 2000.
The first JM to startup will operate. During the first read, a certain amount of
time will be required between bar code detection and cassette insertion.
No specific problems will result and normal reading will be possible.
The first JM to startup will operate. When host names are duplicated on Win-
dows 2000, a warning will be displayed.
If the REGIUS 190's IP and host name are set the same as that for the
REGIUS Console, the Reader will startup, but reading will not be possible. A
"04214" error will occur.
If the REGIUS Console's IP and host name are set the same as that for the
Reader, the REGIUS Console will startup but a "04214" error will occur upon
insertion of a cassette into the Reader and an application error will also
occur for the REGIUS Console.
If the Reader's IP and host name are set the same as that for the JM, a
24000 network error will occur.
If the JM's IP and host name are set the same as that for the Reader, neither
the Reader nor the application will be able to startup.
If the REGIUS Console's IP and host name are set the same as that for the
JM, the REGIUS Console will not startup.
Even if the JM's IP and host name are set the same as that for the REGIUS
Console, normal reading will be possible.