4.2 Troubleshooting for Mechanical Troubles

4.2.1

Caution for Receiver Sensor Plunger

When attempting to remove the back plate
Caution
from the cassette magnetizing plate, always
couple the back plate with the front plate
first, then remove the cassette from the mag-
netizing plate. Removing the back plate
alone after removing the front plate may
cause the edge of the back plate to hit the
sensor plunger of the receiver, and may dam-
age the sensor plunger.