Konica Minolta REGIUS 190 Service Manual page 31

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

1.1 Principal Specifications
Exposure cycle time
The time between
insertion and removal of
the cassette.
Processing Capacity
Sampling Pitch
Max. Reading Size
(mm)
Max. Number of Pixels
QR Value
Dynamic Range (read)
Digital Density Steps
4
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Approximately 40 seconds maximum
(for an image size of 14" x 14" and at 175ı)
90 or more images per hour (for an image size of 14" x 14" and at 175ı)
3 types, i.e.,Super High Precision Mode (43.75µm), High Precision mode
(87.5ı) and Normal mode (175ı)
352.0 x 428.5 (14" x 17" cassette)
7080 x 9480
3 types, i.e., Low Sensitivity (QR value = 125), Standard Sensitivity (QR value =
250), and High Sensitivity (QR value = 500)
• A QR value of 200 is defined for 1,535 steps at an exposure of 2.58 x 10
(=1mR) and 12-bit output for the output signal value.
Normal D range (4.0 digits: PMT output current range = 100 nA to 1 mA)
12 bit (4,096 steps)
-7
C/kg