Cause
12) A Reader and REGIUS
Console share the same
host name.
13) A Reader and JM share
the same host name.
Case 4: Operation status where the network cable has been disconnected during operation.
Cause
1) Connection with the JM was interrupted
when in the Ready condition.
2) Connection with the JM was lost immedi-
ately after the last cassette registration.
3) Connection with the JM was lost in Erase
mode while no cassette was inserted
4) If the network is disconnected for five min-
utes after cassette registration
5) If the network is disconnected for five min-
utes after manual registration of cassette.
6) If the network is disconnected for five min-
utes in the Ready status.
7) If the network is disconnected for five min-
utes during shutdown.
8) During initialization after POWERON, the
network is disconnected for five minutes
from when there is one "
the same for up to ten "
Remedy
Even if the REGIUS Console's host name is set for the Reader, normal read-
ing will be possible.
Even if the Reader's host name is set for the REGIUS Console, normal read-
ing will be possible.
Regardless of the setting location, normal reading will be possible.
Remedy
If no operation was being performed, no Reader problems
will result. If the REGIUS Console is not capable of registra-
tion operation, a restart must be carried out.
If an error is displayed on the Reader during operation, it
will be necessary to restart the Reader and the REGIUS
Console.
If a recorded cassette was read during disconnection
The Reader will not perform reading, operation will stop at the
insertion slot, and a "04214" error will occur. Although a restart will
be necessary for all devices, this can be done for the Reader by
simply turning the power Off and then On. A standard restart will
be required for REGIUS Console and JM.
If after disconnection, a non-operating JM is restarted to return
to the reading condition
The reading of registered images will be possible; however, it will
not be possible to register a new cassette for the REGIUS Con-
sole, and therefore, a restart must subsequently be carried out.
The system will return to Ready mode after the end of Erase
mode. If disconnection continues following this, operation will
be the same as for the above-mentioned disconnection in the
Ready condition.
Normal reading will be possible even after five minutes.
Normal reading will be possible even after five minutes.
Normal reading will be possible even after five minutes.
It will be possible to turn the power off regardless of when the
disconnection takes place during a POWER OFF operation.
A 24000 error occurs for the Reader, and if the network cable is
subsequently connected to re-establish connection with the
". (Operation is
JM, the Ready condition will be achieved and normal operation
will be possible.
".)
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01