Konica Minolta REGIUS 190 Service Manual page 133

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

3.3 Verification of Initializing Sequence
5.
Move the Elevator Unit (Cassette Receive Unit) to the HP.
1)
Click "Move to HP" of "Cassette receive Unit".
2)
Check that "Move to Cassette receive Unit...OK" is
shown on "Information : ".
6.
Check the initialization sequence of Cassette receive Unit.
1)
Click "Initialize" of "Cassette receive Unit".
2)
Check that "Cassette receive Unit Initialize...OK" is
shown on "Information : ".
7.
Move the Cassette receive Unit to the transport standby
position.
1)
Click "Transport pos." of "Cassette receive Unit".
2)
Check that "Cassette receive Unit at Transport-
Standby...OK" is shown on "Information : ".
8.
Check the initialization sequence of sub scanning unit.
1)
Click "Initialize" of "Sub scanning".
2)
Check that "Sub scanning Unit Initialize...OK" is shown
on "Information : ".
106
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01