4.4 Error Code
Error
Error name
code
04906
Cassette warp
detected
04907
Cassette peel off NG
04908
Cassette fall off
detected
162
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Error description
The Peel sensor turned on during
execution of either read
opreations or erasure operations.
The Absorption Detect sensor
turned on after peeling of the cas-
sette from the absorption plate
following joining of the front and
back plates.
It was detected that joining of the
front and back plates was not
confirmed at cassette ejection.
Action
1) Open the front cover and remove any
cassettes, and then close the door
and restore.
2) Halt usage of the problematic cas-
sette and return it to the plant for
repair.
3) If the error occurs with different cas-
settes, adjust the peel detection
roller. (p.117)
1) Remove any cassettes, and then
close the door and restore.
2) If the problem reoccurs for the same
cassette, halt usage of this cassette
and return it to the plant for repair.
1) Remove any cassettes, and then
close the door and restore.
2) Check the operation of the device
and examine the cassettes.