Case 3: Horizontal Streaks In Image - Konica Minolta REGIUS 190 Service Manual

Hide thumbs Also See for REGIUS 190:
Table of Contents

Advertisement

4.3 Troubleshooting for Image Quality
4.3.4

Case 3: Horizontal Streaks in Image

Phenomena
Horizontal streak occurring at one location
1
within the image. *
Horizontal lines extend from shielding
2
materials such as lead.
Sudden change in brightness approximately
3
30 mm from the bottom of the image.
Regular, diagonal stripes This can occur as a
4
result of light leaking from the exterior.
140
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Cause/Remedy
There is a possibility that an external signal is being
applied during reading.
Ensure that the device is protected from external influ-
ences during reading, and perform exposure again.
This can happen when the lead or shielding material is
in a blank section.
This phenomenon happens only under very specific
conditions, and it can be resolved by moving the shield-
ing material as far as possible from the subject or by
restricting the exposure field.
If this continues to occur repeatedly, please contact the
plant or Development.
Return the plate with warping or peeling to the REGIUS
Console.
If the problem repeats with different plates, adjust the
peel detector roller.(p.117)
Plate warping or peeling can lead to contact between
the plate and the detection roller.
Light may be leaking in the device.
Reassemble the exterior covering or darken the room,
and then take a solid image to confirm quality