If the back panel attaches normally, the absorption sensor
turns on.
The subscan motor starts to operate and the absorption
plate moves upward.
If one of the peel sensors turns on, it will indicate that the
plate condition is not normal; in such a case, the upward
motion is halted and the mechanism returns to the subscan
home position.
Several seconds after the V-SYNC sensor turns on, the laser
also turns on and image reading operations start.
When the image reading has been completed, the absorp-
tion plate moves down and simultaneously the erasing
lamps turn on.
The cassette is fed to the ejection shutter in the reverse
order to that described above.
The Cassette receive Unit moves up to the standby position.
The actual stop position will depend on the size of the cas-
sette.
The tip of the cassette comes into contact with the ejection
roller.