4.4 Error Code
Code
Medium Classifi-
cation
Internal error
12205
Internal error
12206
12210
Internal error
12211
Internal error
Main control error
13000
13001
Main control error
Main control error
13030
Main control error
13031
Signal process error
14000
Signal process error
14001
Signal process error
14002
14003
Signal process error
164
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Detailed Clas-
sification
Network com-
Net Command Tim-
munication
eout
Network com-
Transfer End Time-
munication
out
Network com-
Resent counts
munication
Overflow
Network com-
CRC Error counts
munication
Over- flow
H sync timeout
Polygon Timeout
DMA Timeout
SCOM Timeout
Gain Overflow
Gain Underflow
Offset Overflow
Offset Underflow
Countermeasures, etc.
Open and close the door and then
restore. If restoration is not possible,
restart.
Open and close the door and then
restore. If restoration is not possible,
restart.
Open and close the door and then
restore. If restoration is not possible,
restart.
Open and close the door and then
restore. If restoration is not possible,
restart.
No H-SYNC signal is received. Con-
firm H-sync signal . Confirm whether
the laser is ON.
"2.6.5 Checking H-
sync Signal, p.80"
Confirm whether the laser is on. If so,
a polygon is abnormal.
Open and close the door and then
restore. If restoration is not possible,
restart.
Open and close the door and then
restore. If restoration is not possible,
restart.
The gain value has been incorrectly
set. Carry out fixed-current reading
and confirm the operation of the
ADB2. If the fixed current is correct,
carry out a restart. If a problem is
identified, replace the ADB2.
The gain value has been incorrectly
set. Carry out fixed-current reading
and confirm the operation of the
ADB2. If the fixed current is correct,
carry out a restart. If a problem is
identified, replace the ADB2.
The offset value has been incorrectly
set. Carry out fixed-current reading
and confirm the operation of the
ADB2. If the fixed current is correct,
carry out a restart. If a problem is
identified, replace the ADB2.
The offset value has been incorrectly
set. Carry out fixed-current reading
and confirm the operation of the
ADB2. If the fixed current is correct,
carry out a restart. If a problem is
identified, replace the ADB2.