1)
Click [DC motors] tab.
2)
Click [Forward High] of "Insertion Roller".
3)
Wait for 1 second or longer, click [Stop] of "Insertion
Roller".
4)
Click "Others" tab.
5)
Click [Start Read] of "Bar code".
6)
Check that "Reading Bar Code ..." is shown on "Infor-
mation: " window, and the value read in is shown.
6.
The cassette is then drawn into the device, and sand-
wiched by justification mechanism from right and left.
1)
Click [DC motors] tab.
2)
Click [Open] of "Insertion slot shutter".
3)
Click "Stepping motors" tab.
4)
Click [Receive pos.] of "Cassette receive Unit".
5)
Check that "Cassette receive Unit at Receive Position
...OK" is shown on "Information : " window.
Justification Sen.-L (S3) /
Justification Sen.-R (S4)
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.4 Verification of Reading Operation
Justification Motor (PM2)
Cassette Receive Sen. (S2)
Elevator Motor (PM1)