2.7 Electrical Unit
5.
Remove 4 fixing screws, and unplug board connector
(CN3). Then remove the SCB2.
6.
Carry out reassembly in the reverse order to that described
above.
Check that the jumper pin (CN015) is short-circuited.
(other jumper pins should be open)
When the CF card is changed at the same time.
The new CF card contains the default network setting. There-
fore, it may not be possible to communicate with the REGIUS
Console which was changed in setting at the time of installa-
tion. (because of different host name, IP address, subnet set-
ting). In this case, set the network first referring to
Replacement Procedures of CF Card,
nections.
After completing the network connection, make sure that the
data of CF card that has been backed up in the REGIUS Con-
sole is restored onto the new CF card.
I m p or t a n t
Do not use the CF card by copying the
previous data. Always use the CF card
delivered from the factory, which has the
software installed. SCB2 operates on
Linux OS, and its file system employs a
file system which is durable against the
breaker-OFF.
("A.3
p.187"), then make con-
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Fixing Screws
SCB2
CN001
Fixing Screws
CN210
85