If the start up was unsuccessful;
Phenomena
"READY" will not appear on the LCD even
after successfully completing initialization of
the REGIUS 190.
Though "READY" is shown on the LCD
panel of the REGIUS 190, "Starting ..." on
the "Information" window does not change
to screened text.
"COM1 OPEN" is not shown on the "Main"
window, and a message "COM Port not
available" is shown on the pop up.
Either the COM1 port of the mainte-
nance PC is disabled or occupied by
other application.
Action
1) Check that he dIP SW setting on the MCB2.
2) Check that the cables are properly connected.
3) Restart the "dummy SBC2 program", and check that the
"Starting..." is on the "Information" window is shown in black.
4) Restart the REGIUS 190.
1) Check that he dIP SW setting on the MCB2.
2) Restart the REGIUS 190.
1) Select "Setup" --> "Communication Port" in the main window
menu.
2) Select an available port in the combo box for COM ports.
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.2 Set Up of dummy SBC2 program