GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-3-2
Monitor Cable Assy (FRU No. 101)
Purpose: This is a description on how to remove and replace the Monitor Assy.
8-3-2-1
Tools
•
Common pillips screwdrivers
•
Cutter
8-3-2-2
Needed Manpower
•
1person, 5 minutes + travel
8-3-2-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
•
Maneuver Control Console to a suitable position for removing the monitor.
8-3-2-4
Removal procedure
1.) Remove the OP Rear Cover. Refer to section 8-5-10 on page 58.
2.) Remove the Monitor Covers. For more information, refer to section 8-3-3 on page 8.
3.) Disconnect two connectors (1-2) and cut the tie wrap in the location (3). Refer to Figure 8-7.
Connector 1
Connector 2
8 - 6
7
Figure 8-7 Disconnect Connectors and Cut the tiewrap
Section 8-3 - Monitor
LOGIQ™ 5 PRO S
ERVICE
Tie wrap
M
ANUAL