Setting Institutional Defaults (Sample)

O
M
N-600x Service Manual
XI
AX
BATTERY Softkey
The BATTRY softkey places the monitor into shelf-mode while the monitor is in
storage in order to preserve the life of the battery. This feature will be available in
future releases of the O
XI
Power-up default values may be changed to institutional power-up default values.
Set the desired limits in the normal operation mode and set the institutional
defaults in the monitor's service mode. The following default values may be set:
Table 5: Institutional Default Settings
Parameter
ALARM SILENCE Duration
Alarms
Alarm Volume
Sensor Event Record Type
Nurse Call Priority RS-232
Pulse Beep Volume
Pulse Rate Upper Alarm Limit
Pulse Rate Lower Alarm Limit
SatSeconds
SatSeconds
Allow
Sensor Adjust Enable
Data Port Baud Rate
Data Port Mode
SpO
Upper Alarm Limit
2
SpO
Lower Alarm Limit
2
Power-On Settings and Service Functions
M
N-600x monitor.
AX
Setting
OFF, 30, 60, 90, 120 seconds
(Service Mode settings only)
Allow Off - Yes/No
1 to 10
(min 45dB, max > 85dB)
%SpO
, %SpO
2
normally high, normally low
0 to 10
Lower Limit plus 1 to 250 BPM
20 BPM to Upper Limit minus 1
OFF, 10, 25, 50, 100 seconds
Yes/No
Yes/No
2400, 9600, 19200
ASCII, OXINET, CLINICAL, GRAPH,
PHILIPS, SPACELABS, GE
MARQUETTE, DATEX OHMEDA AS/3
monitors. Available selections depend on
the software installed in your monitor.
Lower Limit plus 1 to 100%
Upper Limit minus 1 to 85%
+ BPM, Default
2
61