The Datex-Ohmeda AS/3 monitor must be configured for communications with
the Nellcor N-200 monitor in order to communicate with the N-600x monitor.
Refer to the AS/3 Operator's Manual for instructions on configuring the AS/3
monitor. The N-600x sends SpO
AS3 monitor. The RS-232 hardwire interface cable has a DB-15 connector for the
monitor and the applicable connector for the Datex monitor. Nellcor cable part
number 902255 is recommended for this interface. Corrupt data is indicated by a
Communications Error displayed on the Datex monitor. When the monitor is in
the Datex mode of operation the interface baud rate is automatically set to 2,400
bits per second.
WARNING: Do not silence or decrease the volume of the N-600x audible
alarm if patient safety could be compromised.
The Datex-Ohmeda AS/3 monitor does not indicate audible or visual alarms for
equipment interfaced. The monitor must be able to sound an audible alarm in
order to maintain patient safety.
Note: The parameters setup for the Datex-Ohmeda bedside monitor interface
may be saved as institutional default settings. See WARNING: Audible
alarms should not be silenced if patient safety could be compromised. on
page 60. Use care when performing this procedure since all settings are
saved as institutional default settings.
Data Port Interface Protocol
, pulse rate, and alarm status data to the Datex