Data Port Interface Protocol

SpaceLabs Communications

GE Marquette Communications

134
The N-600x sends %SpO
monitor. The SpaceLabs monitor requires a Universal FlexPort
module to interface with the N-600x pulse oximeter. Corrupt data is indicated by a
Communications Error displayed on the SpaceLabs monitor. When the monitor is
in the SpaceLabs mode of operation the interface baud rate is automatically set to
9,600 bits per second.
WARNING: Do not silence or decrease the volume of the N-600x audible
alarm if patient safety could be compromised.
The SpaceLabs monitor provides both audible and visual alarm indications for
equipment interfaced through the Universal FlexPort
Silencing the monitor alarms also silences the SpaceLabs monitor alarms. The
monitors must be able to sound an audible alarm in order to maintain patient
safety.
Note: The parameters setup for the SpaceLabs bedside monitor interface may be
saved as institutional default settings. See WARNING: Audible alarms
should not be silenced if patient safety could be compromised. on page 60.
These settings will be saved as institutional default settings.
The monitor sends %SpO
monitor. The Marquette monitor requires an Octanet
interface with the N-600x pulse oximeter. The interface module comes with an
interface cable, GE Marquette part number 417961-033, that connects to the
Nellcor interface cable. The RS-232 hardwire interface cable has a DB-15
connector for the monitor and the applicable connector for the Marquette
Octanet
TM
interface module cable. Nellcor cable part number 902254 is
recommended for this interface. Corrupt data is indicated by a Communications
Error displayed on the GE Marquette monitor. When the N-600x is in the GE
Marquette mode of operation the interface baud rate is automatically set to 9,600
bits per second. The GE Marquette monitor only sounds audible alarms for
equipment interfaced through the Octanet
monitor audible alarm has no effect on the GE Marquette monitor sounding an
alarm.
Note: The parameters setup for the GE Marquette bedside monitor interface
may be saved as institutional default settings. See WARNING: Audible
alarms should not be silenced if patient safety could be compromised. on
page 60. Use care when performing this procedure since all settings will
be saved as institutional default settings.
, pulse rate, and alarm status data to the SpaceLabs
2
, pulse rate, and alarm status data to the GE Marquette
2
TM
interface module. Silencing the
TM
interface
TM
interface module.
TM
interface module to