Nellcor OxiMax N-600x Service Manual page 149

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

O
M
N-600x Service Manual
XI
AX
Patient Data
N-600x
VERSION X.X.X.X.
ADULT
TIME
%SpO2
01-JAN-06
14:00:05
100
Patient data and the operating status of the unit are highlighted in the display
above. Parameter values, at the time of the printout, are displayed directly beneath
the heading for each parameter. In this example the %SpO
rate (BPM) is 190 beats per minute. The asterisk (*) next to the 190 indicates that
190 beats per minute is outside of the alarm limits, indicated in the top row, for
pulse rate. If no data for a parameter is available, three dashes [- - -] display on the
printout.
Pulse Amplitude (PA) can range from 0 to 254. There are no alarm parameters for
this value. It can be used for trending information and is an indication of a change
in pulse volume, pulse strength, or circulation.
Operating Status
N-600x VERSION X.X.X.X CRC: XXXX SpO2 Limit: 70-100%
ADULT
TIME
01-JAN-06 14:00:05
The Status column indicates alarm conditions and operating status of the N-600x.
In this example, PH refers to Pulse High. The status codes are listed in Table 22.
As many as four codes can be displayed simultaneously in the Status column.
Table 22: Operating Status Codes
Code
AO
AS
BU
LB
LM
LP
MO
PH
PL
CRC: XXXX
SpO2 Limit: 70-100%
0SAT-S
SPO2 RESP MODE: NORMAL
BPM
PA
Status
*190
50
0SAT-S
SPO2 RESP MODE: NORMAL
%SpO2
BPM
PA
100
*190
50
Meaning
Alarm Off
Alarm Silence
Battery in Use
Low Battery
Loss of Pulse with Interference
Loss of Pulse
Interference Detected
Pulse Rate Upper Limit Alarm
Pulse Rate Lower Limit Alarm
Data Port Interface Protocol
PR Limit: 60-160BPM
is 100, and the pulse
2
PR Limit: 60-160BPM
Status
PH
143