Analog Calibration Setup

Philips Communications

O
M
N-600x Service Manual
XI
AX
Analog calibration signals are provided to adjust a recorder to the output of the
instrument. Selectable calibration signals are +1.0 VDC, 0.0 VDC, and Step. For
more information on the analog signals see Analog Output on page 146.
To adjust the analog calibration signal:
1. Turn on the monitor by pressing the ON/STANDBY button.
2. Press the SETUP softkey.
3. Press the NEXT softkey three (3) consecutive times.
4. Press the ANALOG softkey.
0 VOLT
1 VOLT
5. Press the 0 VOLT, 1 VOLT, or STEP softkey as required.
6. Press the BACK softkey.
The N-600x sends %SpO
The Philips monitor requires a Philips VueLink
(A05 option) to interface with the N-600x pulse oximeter.
The RS-232 hardware interface cable has a DB-15 connector for the N-600x and
the applicable connector for the Philips monitor. The Nellcor cable part number
902256 is recommended for this interface.
A blank screen on the Philips monitor indicates corrupt data. The Philips monitor
detects corrupt data in less than 100 milliseconds. When the N-600x is in the
Philips mode of operation, the interface baud rate is automatically set to 19,200
bits per second (bps).
%SP02
100
BPM
BACK
STEP
, pulse rate, and alarm status data to the Philips monitor.
2
TM
Data Port Interface Protocol
---
---
Aux Plus B interface module
133