In the INSTAT printout below, line one is for instrument type, software revision
level, type of printout, and alarm parameter settings. The second line contains the
column headings. A trend point is recorded for every four seconds of instrument
operation. Up to 48 hours of instrument operation data can be recorded.
N-600x VERSION X.X.X.X
Instat
ADULT
0SAT-S
TIME
Trend 01
%SpO2 BPM
01-JAN-06 14:00:00
- - -
01-JAN-06 14:00:05
- - -
01-JAN-06 14:00:10
100
01-JAN-06 14:00:15
100
N-600x VERSION X.X.X.X
Instat SpO2 Limit: 80-100%
ADULT
0SAT-S
TIME TREND 02
%SpO2 BPM
01-JAN-06 14:24:24
79*
01-JAN-06
14:24:29
79*
01-JAN-06
14: 24:29
0*
N-600x VERSION X.X.X.X
Instat SpO2 Limit: 80-100%
ADULT
0SAT-S
TIME TREND 03
%SpO2 BPM
01-JAN-06
7:13:02
99
01-JAN-06
7:13:07
99
01-JAN-06
7:13:12
99
01-JAN-06
7:13:17
99
01-JAN-06
7:13:22
99
01-JAN-06
7:13:27
99
01-JAN-06
7:13:32
99
Output Complete
If the final line on the printout shows "Output Complete," then the data has been
successfully transmitted with no corruption. If there is no "Output Complete" line
printed, the data should be considered invalid.
INFO Softkey
The INFO softkey is used by Nellcor's Customer Service Engineering only.
Pressing the INFO softkey produces a single line printout of instrument
information as illustrated below. The data presented in the printout from left to
right, is the instrument type (N-600x), software version level, type of printout
(INFO), CRC (Cyclic Redundancy Check) number, and ratio of current operating
time to total operating time (the ratio itself has no units of measure).
N-600x Version XXXXXX INFO CRC:XXXX SEC: 123456789/987654321
Power-On Settings and Service Functions
SpO2 Limit: 30-100%
PR Limit: 100-180BPM
SPO2 RESP MODE: NORMAL
PA
Status
- - -
- - -
SD
- - -
- - -
PS
120
220
120
220
PR Limit: 60-180BPM
SPO2 RESP MODE: NORMAL
PA
Status
58*
220
PS SL PL
57*
220
PS SL PL
0*
- - - PS LP SL PL
PR Limit: 60-180BPM
SPO2 RESP MODE: NORMAL
PA
Status
132*
220
PH
132*
220
PH
132*
220
PH
132*
220
PH
132*
220
PH
132*
220
PH
132*
220
PH
57

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents