Power-On Settings and Service Functions
62
To set institutional defaults:
1. Disconnect the O
M
XI
Note: If the O
M
sensor is not disconnected, the only softkeys on the screen
XI
AX
will be PARAM and EXIT.
2. Set desired parameters to the institutional values. Refer to the O
Operator's Manual for parameter values.
3. Simultaneously press the LIGHT softkey and the HELP/CONTRAST button
until the softkey headings appear as shown below.
PARAM
4. Press the PARAM softkey.
RESET
5. Press the SAVE softkey.
SAVE DEFAULTS?
6. Press the YES softkey. The monitor sounds three beeps indicating the defaults
have been saved.