Service Functions

Accessing the Service Functions

O
M
N-600x Service Manual
XI
AX
Service functions can be used to select institutional defaults and to access
information about the patient or instrument. Only a Nellcor Customer Service
Technician should access many of the items available through the service
functions.
All service functions are accessible when the DOC-10 pulse oximetry cable is
disconnected from the monitor. Disconnect the O
pulse oximetry cable; or, disconnect the DOC-10 pulse oximetry cable from the
instrument.
To access the service functions:
1. Turn on the monitor by pressing the ON/STANDBY button.
2. Wait for monitor power-on self-test to complete.
3. Simultaneously press and hold the LIGHT softkey and the
HELP/CONTRAST button until the service softkeys display.
PARAM
Note: The service function is only accessible from the Main menu display. See
PARAM (Service Menu) on page 52.
Note: If the above step is performed with a pulse oximetry cable connected,
only the PARAM and EXIT softkeys appear on the screen.
Power-On Settings and Service Functions
XI
O
M
XI
AX
%SP02
BPM
PRINT
NEXT
EXIT
M
sensor from the DOC-10
AX
N-600x
---
---
?
51