Performance Verification
38
3. Turn on the monitor by pressing the ON/STANDBY button. After the monitor
completes the self-check, the following screen displays:
Active audio alarm.
Flashing %SpO
2
BPM indication between 57 and 63 inclusive.
Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual
amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
4. Press the SRC-MAX MODULATION selection button. The SRC-MAX
Modulation
LED lights. The monitor pulse amplitude waveform initially
increases in amplitude and then stabilize at P-T-P amplitude of approximately
1-inch. The following screen displays:
Active audio alarm.
Flashing %SpO
2
BPM indication between 57 and 63 inclusive.
Pulse waveform of approximately 1-inch P-T-P amplitude. Actual
amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
O
%SP02
BPM
LIMITS
TREND
SETUP
LIGHT
indication between 73 and 77 inclusive.
O
XI
%SP02
BPM
LIMITS
TREND
SETUP
LIGHT
indication between 73 and 77 inclusive.
M
N-600x
XI
AX
75
60
?
M
N-600x
AX
75
60
?