Performance Verification
34
3. Turn on the monitor by pressing the ON/STANDBY button. After the monitor
completes its self-check, the following screen displays:
Active audio alarm.
Flashing %SpO
2
BPM indication between 57 and 63 inclusive.
Pulse waveform of approximately 1/2-inch peak to peak (P-T-P)
amplitude. Actual amplitude may vary but will be a reference for low
pulse amplitude/low light patients.
4. Press the SRC-MAX PULSE RATE selection button. The SRC-MAX PULSE
RATE 200 LED lights. The monitor BPM increases and stabilizes to a value
between 197 to 203 BPM inclusive. The following screen displays:
Active audio alarm.
Flashing %SpO
2
Flashing BPM indication between 197 and 203.
Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual
amplitude may vary but references low pulse amplitude/low light
patients.
O
%SP02
BPM
LIMITS
TREND
SETUP
LIGHT
indication between 73 and 77 inclusive.
%SP02
BPM
LIMITS
TREND
SETUP
LIGHT
indication between 73 and 77 inclusive.
M
N-600x
XI
AX
75
60
?
N-600x
O
M
XI
AX
75
200
?