Nellcor OxiMax N-600x Service Manual page 65

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

O
M
N-600x Service Manual
XI
AX
SELECT Softkey
The SELECT softkey is used to select which function of the audible alarm
changes. A box can be cycled between ALLOW OFF and OFF REMINDER.
Use the following procedure to select and set the monitor's ALLOW OFF and
OFF REMINDER:
1. Disconnect the O
M
XI
Note: If the O
M
sensor is connected, the only softkeys on the monitor's
XI
AX
screen display are PARAM and EXIT.
2. Simultaneously press the LIGHT softkey and the HELP/CONTRAST button
until the menu bar changes to the softkey headings shown below.
PARAM
3. Press the NEXT softkey.
DOWNLD
4. Press the ALARMS softkey.
ALLOW OFF?
OFF REMINDER?
SELECT
Power-On Settings and Service Functions
sensor from the monitor.
AX
O
M
XI
AX
%SP02
BPM
PRINT
NEXT
EXIT
O
XI
%SP02
BPM
NEXT
EXIT
ALARMS
O
M
XI
ALARMS
%SP02
YES
BPM
YES
BACK
N-600x
---
---
?
M
N-600x
AX
---
---
?
N-600x
AX
---
---
?
59