Param Softkey Menu

O
M
N-600x Service Manual
XI
AX
When the PARAM softkey is pressed, the function of the softkeys changes as
shown below. These options can be accessed with the pulse oximetry cable
connected to the instrument.
RESET
SAVE
Reset Softkey
The RESET softkey can be used if any settings stored in memory have been
changed from factory default values. Pressing YES sounds three tones and returns
the settings to factory default values. Pressing NO stores the instrument settings
into memory.
Save Softkey
When adjustable values are changed from factory default, the SAVE softkey can
be used to preserve the settings as institutional power-on default values. Pressing
YES stores the current settings in memory. The instrument sounds three tones
indicating that the changes have been saved as power-on default values. The new
saved values continue to be used through power-on and off cycles until they are
changed and saved again, or until they are reset. Pressing NO restores the original
default values.
Note: An invalid tone indicates a parameter value cannot be saved as a
power-on default. See Adjusting the Power-On Settings on page 47.
Along with the invalid tone, a message displays indicating which
parameter could not be saved as a power-on default.
Sensor Softkey
The SENSOR softkey enables/disables the Sensor Event Record function.
Power-On Settings and Service Functions
O
M
N-600x
XI
AX
---
%SP02
---
BPM
BACK
SENSOR
?
53